What it is for
A pH meter measures the voltage between a glass probe sensitive to H+ ions and a reference electrode. It reads to 0.01 pH, which matters for demanding ferments (cured meats, cheeses, vinegars).
When you need it
Cured meats (the pH drop in the first days), aged cheeses, professional fermentation. Also to check what your strips tell you.
- Precision
- ±0.01–0.02 pH right after calibration. Calibrate before every session that matters.
What you need
- Digital pH meterResolution 0.01 pH, interchangeable probe, 2-3 point calibration
- Buffer solutions pH 4.01 and pH 7.00Lab supply shops online (€3–8 per sachet)
- 3M KCl storage solutionLab supply shops online
How to do it
- Bring the buffers to room temperaturePour pH 4.01 and 7.00 into two small cups and let them reach 20–25 °C.Why. Cheap meters without ATC need the sample at room temperature.
- Calibrate first point at pH 7.00Rinse the probe with distilled water, blot it gently, dip it in pH 7.00, press CAL and wait 30 s.Careful. Never rub the glass bulb: it scratches the membrane and ruins the meter.
- Calibrate second point at pH 4.01Rinse, blot, dip it in 4.01 and press CAL again. The meter works out the slope by itself.
- Measure the sampleRinse between samples. Dip it in the ferment, wait 30–60 s, and read once the display is stable.
- Store the probe properlyNever let it dry: keep it in the cap with 3M KCl. If it dried out, rehydrate it for 24 hours before use.Why. A dry probe may still work, but it takes minutes instead of seconds to respond and is less accurate.
Common mistakes
Not calibrating, or calibrating rarely
- What happens
- Drift of up to ±0.3 pH: wrong decisions where it matters
- How to avoid it
- Calibrate every session, or at least weekly
Storing the probe in distilled water
- What happens
- Pure water leaches the membrane: its life drops from years to months
- How to avoid it
- 3M KCl or a dedicated storage solution
Rubbing it with rough paper
- What happens
- Micro-scratches on the membrane → slow, imprecise readings
- How to avoid it
- Rinse, then blot gently with a soft tissue
Safety
- Buffer solutions can irritate eyes and skin: rinse if they touch you
The why, in short
The glass bulb holds an internal solution of known pH. In the sample it develops a voltage proportional to the pH difference (Nernst: 59 mV per unit at 25 °C).
Written by the Fermentation Lab team, the same text as the app's Library. Found a mistake? Write to us: fermentationlabapp@gmail.com